EC-20 cond

All features at a glance

• Display Conductivity and
  Temperature simultaneously
• Automatic calibration 1 to 3
  points between 4 conductivity
  standard values
• Manual calibration on 1 point
  (User defined)
• Large Display
  (backlight for EC-25 cond)
  with indication of standards used
  for calibration and stability
• Automatic and manual compensation
  of temperature with reference
  value 15...30 °C
• Waterproof IP57 protection
• Strong carrying case with all
  accessories to be used
  as a small portable laboratory


EC-25 cond

additional for EC-25 cond

• TDS measurement with adjustable factor
• GLP
• Memory recall of last calibration with control of cell constant,date and time
• CAL DUE : set of calibration frequency for high quality measurements
• Data logger, manual and automatic with 500 memory values with date and time
• USB port to download data and power by Computer or power supply (included)
• Software PC-Link to download data (included)


Conductivity Portable Meter EC-20 cond • EC-25 cond

An elegant and ergonomic case, with a practical soft rubber andIP57 protection, has inside the best of technology. A large display (backlight for EC-25 COND) shows all information that you need about the measurement, cell‘s life and instrument‘s memory. All the functional parameters of connected cell, during measurement or calibration, are constantly checked and inform theoperator for easy trouble-shooting.

Only 4 keys for COND 7 (6 for EC-25 cond) allow the user to control all functions in an intuitive and easy way. A practical USB port (EC-25-cond), covered to guarantee waterproof protection, allow to download data (with date of last calibration done) or power the instruments by Computer or with power supply 220V/USB included.

Technical data

Measuring Range (cond)

Resolution (cond)
Calibration Standards

Calibration points
Cell constant selectable
(cm-1)
Reference Temp. (°C)
Temperature compens.(%/°C)
Measuring Range (°C)
Resolution (°C)
TDS-Measuring Range (g/l)
TDS-Factor
Salinity (ppt)
Memory
Timer
GLP
Interface
Display
IP-Protection
Power
Weight (g)
Dimension (mm)
EC-20 cond

0,00 - 20,00 - 200,0 -
2000 µS 2,00 - 20,00 - 200,0 mS 0,01/0,1/1 µS 0,01/0,1 mS
84, 1413 µS, 12.88,
111.8 mS, User
1...4
0,1 – 1 – 10

15...30
0,00...10,00
0...100,0
± 0,5
0...100
0,4...1
No
No
No
No
No
LCD
IP 57
Battery
295
86 x 196 x 33 (WxDxH)
EC-25 cond

0,00 - 20,00 - 200,0 - 2000 µS 2,00 - 20,00 - 200,0 mS
0,01/0,1/1 μS 0,01/0,1 mS
84, 1413 µS, 12.88,
111.8 mS, User
1...4
0,1 – 1 – 10

15...30
0,00...10,00
-10...110
± 0,5
0...100
0,4...1
0...50
500 Data
Yes
Yes
Yes
LCD, 3 color backlit display
IP 57
Battery
295
86 x 196 x 33 (WxDxH)

Various sets are available:

EC-20 cond 0
Kit without cell, with standard calibration solutions,carrying case and accessories

EC-20 cond 1
Complete kit include cond. Cell 2301T with ATC, cable length 1m and BNC plug, standard calibration solutions, carrying case and accessories

EC-25 cond 0
Kit without cell, with standard calibration solutions, power supply, software PC-Link, USB cable, carrying case with accessories

EC-25 cond 1
Complete kit include cond. Cell 2301T with ATC, cable length 1m and BNC plug, Standard calibration solutions, power supply, software PC-Link to download data, USB cable, carrying case with accessories



Downloads (PDF)


Elektrodes

Elektrochemistry